Please use this identifier to cite or link to this item: http://dspace.aiub.edu:8080/jspui/handle/123456789/1663
Title: Ventilation mapping of chest using Focused Impedance Method (FIM)
Authors: Kadir, Abdul Kadir
Ferdous, Humayra
Baig, Tanvir
Rabbani, K Siddique
Keywords: Focused Impedance Method (FIM), Lung Impedance,
Issue Date: 8-Apr-2010
Publisher: Journal of Physics: Conference Series
Citation: 14
Series/Report no.: 224 (2010);012031
Abstract: Focused Impedance Method (FIM) provides an opportunity for localized impedance measurement down to reasonable depths within the body using surface electrodes, and has a potential application in localized lung ventilation study. This however needs assessment of normal values for healthy individuals. In this study, localized ventilation maps in terms of electrical impedance in a matrix formation around the thorax, both from the front and the back, were obtained from two normal male subjects using a modified configuration of FIM. For this the focused impedance values at full inspiration and full expiration were measured and the percentage difference with respect to the latter was used. Some of the measured values would have artefacts due to movements of the heart and the diaphragm in the relevant anatomical positions which needs to be considered with due care in any interpretation.
URI: http://dspace.aiub.edu:8080/jspui/handle/123456789/1663
Appears in Collections:Publication: Journal

Files in This Item:
File Description SizeFormat 
Paper _2010.docx70.68 kBMicrosoft Word XMLView/Open